防止生物危机转子 AF-2730, AT-2730M, AF-5004CS |
|
In the case of handling
bio hazardous experiments, such as Bovine Spongiform Encephalopathy
(BSE) tests, researchers must ensure that the rotors are
biohazard-proof. If the tubes were broken by chance during
the centrifugation, samples would become aerosol and spread
around, increasing the biohazard danger against the researchers
tremendously.
Biohazard-proof rotors, such as AF-2730, AT-2730M, and
AF-5004CS, are ensured to maintain the aerosol and sample
liquid between the rotor and its lid, certified
with CAMR standard test procedures. AF-2730
and AT-2730M accept 30 x 2mL tubes over 21,000Xg, while
AF-5004CS accepts 4 x 50mL TC tubes over 14,000Xg. In
addition, AF-2730 and AT-2730M, including the rotor and
its lid, are autoclavable
to eliminate bio hazardous contamination such as prion. |
|
|
|
Fixed Angle
Rotor - AF-2730 |
|
|
AF-2730 accepts
30 x 2mL and 14 x 0.5 and 0.2mL tubes. |
|
The rotor and its
lid are autoclavable 50 times at 121°C or 134°C
to eliminate bio hazardous contamination. |
|
The maximum rpm
and g-force are 14,000rpm and 21,036Xg respectively. |
|
AF-2730 is available
with 3400, 3700, 3740, 3780 and 6200. |
|
|
|
|
Fixed Angle
Rotor - AT-2730M |
|
|
AT-2730M accepts
30 x 2mL and 14 x 0.5 and 0.2mL tubes. |
|
The rotor and its
lid are autoclavable 50 times at 121°C or 134°C
to eliminate bio hazardous contamination. |
|
The maximum rpm
and g-force are 14,000rpm and 21,036Xg respectively. |
|
AT-2730M is available
with 5922. |
|
|
|
|
Fixed Angle
Rotor - AF-5004CS |
|
|
AF-5004CS accepts
4 x 50mL TC tubes |
|
The rotor and its
lid are autoclavable 20 times at 121°C or 134°C
to eliminate bio hazardous contamination. |
|
The maximum rpm
and g-force are 12,000rpm and 14,970Xg respectively. |
|
AF-5004CS is available