Swinging bucket rotor for the 7700 model |
Rotor
Name |
Nominal
Capacity |
Tube
Size (mm)
diam. x length |
Maximum
Speed
(rpm) |
Maximum
RCF
(xg) |
Bucket
Code No. |
Tube
Rack Adaptor
Code No. |
Swinging
Bucket Rotor |
RS-410G
(*1,4) |
48
x 2mL |
10.7
x 42 |
5,000 |
4,160 |
Bucket
included |
S17008 |
48
x 5mL |
12.5
x 75 |
5,000 |
4,160 |
Bucket
included |
S17006 |
48
x 7mL |
12-12.5
x 100-112 |
5,000 |
4,160 |
Bucket
included |
S17007 |
28
x 7mL blood collecting tube |
12.5-13
x 75-106 |
5,000 |
4,160 |
Bucket
included |
S17027 |
12
x 15mL Conical (TC)(*2) |
16.5
x 120.6 |
5,000 |
4,160 |
Bucket
included |
S17019 |
12
x 15mL Conical (TC)(*2) |
16.5
x 120.6 |
5,000 |
4,470(*5) |
S13130 |
- |
28
x 10mL blood collecting tube/15mL |
15.5-17
x 90-110 |
5,000 |
4,160 |
Bucket
included |
S17015 |
4
x 50mL |
34
x 110 |
5,000 |
4,160 |
Bucket
included |
S17050 |
4
x 50mL |
27-36.5
x 75-110 |
5,000 |
4,160 |
Bucket
included |
055-5670 |
4
x 50mL Conical (TC) |
29.5
x 117 |
5,000 |
4,160 |
Bucket
included |
S17049 |
4
x 100mL |
44-45
x 110 |
5,000 |
4,160 |
Bucket
included |
S17100 |
4
x 200mL |
56.5
x 104 |
5,000 |
4,160 |
Bucket
included |
- |
返回页首 |
RS-720G |
80
x 2mL |
9.5-11
x 36-42 |
3,500 |
2,290(*5) |
053-5820 |
055-4800 |
96
x 1.8mL Sample cup |
17
x 38 |
3,500 |
2,040(*5) |
053-5820 |
055-4940 |
96
x RIA tube |
10-12
x 70-95 |
3,500 |
2,190(*5) |
053-5820 |
055-4730 |
80
x 10mL |
12-14
x 70-95 |
3,500 |
2,260(*5) |
053-5820 |
055-4740 |
48
x 10m blood collecting tube/15mL |
15-17.2
x 75-110 |
3,500 |
2,330 |
053-5820 |
055-4750 |
68
x 7m blood collecting tube |
12.5-13.5
x 75-110 |
3,500 |
2,300 |
053-5840 |
055-0260 |
68
x 10mL blood collecting tube/15mL |
15-17.2
x 75-110 |
3,500 |
2,300 |
053-5840 |
- |
28
x 15mL Conical (TC) |
17
x 121 |
3,500 |
2,300 |
053-5840 |
- |
32
x 15mL Conical (TC) |
17
x 121 |
3,500 |
2,330 |
053-5820 |
055-4780 |
8
x 50mL |
27-36.5
x 75-110 |
3,500 |
2,330 |
053-5820 |
055-4760 |
16
x 50mL Conical (TC) |
30
x 117 |
3,500 |
2,330 |
053-5820 |
055-6070 |
4
x 100mL |
38-46.5
x 75-120 |
3,500 |
2,330 |
053-5820 |
055-4770 |
4
x 175mL Conical (TC)(*3) |
61
x 118 |
3,500 |
2,330 |
053-5820 |
055-4850 |
4
x 200mL Conical (TC)(*3) |
61
x 130 |
3,500 |
2,330 |
053-5820 |
055-4850 |
4
x 225mL Conical (TC)(*3) |
61
x 137 |
3,500 |
2,330 |
053-5820 |
055-4850 |
4
x 250mL |
60-61.8
x 100-130 |
3,500 |
2,380(*5) |
053-5820 |
055-4850 |
RS-730G |
80
x RIA tube |
12
x 75-110 |
4,200 |
3,589 |
S13215 |
S18005 |
48
x 7mL blood collecting tube |
12.5-13
x 75-106 |
4,200 |
3,589 |
S13215 |
S18012 |
48
x 15mL(*7) |
15.5-17
x 96-102 |
4,200 |
3,589 |
S13215 |
S18015 |
16
x 15mL Conical (TC) |
16.5
x 120.6 |
4,200 |
3,589 |
S13215 |
S18016 |
8
x 50mL Conical (TC)(*6) |
29.5
x 117 |
4,200 |
3,589 |
S13215 |
S18052 |
28
x 50mL Conical (TC) |
29.5
x 117 |
4,200(*8) |
3,750(*5) |
S13235 |
- |
12
x Microtiter Plate(*8) |
86(W)
x 130(D) x 49(H) |
4,200 |
3,180(*5) |
S13220 |
- |
4
x Deep-well Plate(*8) |
返回页首 |
RS-3011G
(*9) |
160
x 1.8mL Sample cup |
17
x 38(38) |
3,300 |
2,280 |
053-5310 |
055-4950 |
140
x 6mL/RIA tube |
10-12
x 65-100(95) |
3,300 |
2,150(*5) |
053-5310 |
055-5830 |
100
x 10mL |
12-14.5
x 75-110(105) |
3,300 |
2,280 |
053-5310 |
055-5840 |
80
x 15mL(64) |
14.5-17.2
x 75-110(105) |
3,300 |
2,280 |
053-5310 |
055-5850 |
16
x 15/50mL |
15mL
14.5-17.2 x 75-135(125) |
3,300 |
2,280 |
053-5310 |
055-5860 |
50mL
27-36 x 75-115(110) |
48
x 15mL Conical (TC) |
17
x 121(121) |
3,300 |
2,280 |
053-5310 |
055-4660 |
24
x 50mL Conical (TC) |
30
x 117(117) |
3,300 |
2,280 |
053-5310 |
055-4610 |
8
x 100mL |
38-46.5
x 75-120(116) |
3,300 |
2,280 |
053-5310 |
055-5870 |
4
x 250mL |
60-61.8
x 100-141(136) |
3,300 |
2,280 |
053-5310 |
055-6880 |
4
x 350mL |
65-71
x 100-130(122) |
3,300 |
2,280 |
053-5310 |
055-6890 |
96
x 10mL |
12-14
x 75-110 |
3,300 |
2,240(*5) |
053-5180 |
055-0200 |
72
x 15mL |
14.5-17.2
x 75-110 |
3,300 |
2,280 |
053-5180 |
055-0210 |
16
x 50mL |
27-36
x 75-114 |
3,300 |
2,280 |
053-5180 |
055-0240 |
48
x 15mL Conical (TC) |
17
x 121 |
3,300 |
2,280 |
053-5180 |
055-0220 |
16
x 50mL Conical (TC) |
30
x 117 |
3,300 |
2,280 |
053-5180 |
055-0230 |
4
x 250mL(*3) |
60-61.8
x 100-141 |
3,300 |
2,280 |
053-5180 |
055-4490 |
4
x 350mL |
65-71
x 100-130 |
3,300 |
2,280 |
053-5180 |
055-5880 |
4
x 500mL Conical (TC)(*10) |
96
x 147 |
3,300 |
2,280 |
053-5180 |
431124
(*10) |
4
x 650mL |
84-86
x 100-125 |
3,300 |
2,280 |
053-5180 |
055-5890 |
4
x 750mL |
95-97.5
x 100-135 |
3,300 |
2,280 |
053-5180 |
- |
8
x Microtiter Plate(*4) |
88(W)
x 134(D) x 50(H) |
3,300 |
2,020(*5) |
053-5220 |
- |
4
x Deep-well Plate(*4) |
8
x Microtiter Plate(*4) |
92(W)
x 133(D) x 36(H) |
2,000 |
770(*5) |
053-5170 |
055-5900 |
8
x Tissue-culture Plate(*4) |
2,000 |
770(*5) |
053-5170 |
- |
|
|
(*1) |
RS-410G Rotor can accommodate
sealing caps (code no. S17111) (Optional). |
(*2) |
The sealing cap cannot
be used with the combination of the original bucket and
S17019 tube rack.
The bucket for 12 x 15 conical (TC) tube (code no. 13130)
able to accept the sealing caps. |
(*3) |
The 175mL Conical (TC)
is made by Falcon Corporation (Code No. 352076) and requires
a cushion (Code No. 352090).
The 200mL Conical (TC) is made by Nalge Nunc International
(Code No. 376813) and requires a cushion adapter (Code
No. 377585).
The 225mL Conical (TC) is made by Falcon Corporation (Code
No. 352075) and requires a cushion (Code No. 352090). |
(*4) |
Regarding the delivery
schedule, please contact your local dealer. |
(*5) |
More information for
the display of centrifugal force, please refer the Instruction
Manual. |
(*6) |
FALCON®, SARSTEDT®,
Lab-Tek®, GREINER(White), and IWAKI 50mL conical can
tube can use shield covers. For more information, contact
your local dealer. |
(*7) |
When shield covers are
used, tubes cannot be inserted in the tube holes at the
four corners of the bucket. Accordingly, the number of
tubes for each bucket is reduced by four tubes. (Example:
When 15mL glass tubes are used in bucket S13215, up to
32 tubes can be used). |
(*8) |
If microtiter plates
are used, it is possible to stack them in double or more
on each bucket. However, be sure to test the strength
of those plates prior to centrifugation. The microtiter
plate at the bottom may be crushed. |
(*9) |
The length inside ()
includes the length of shield cover (code no. 027-9210). |
(*10) |
A bottle made by Corning
must be used (Code No. 431123). And also cushion adapter
made by Corning is required (Code No. 431124). |
返回页首 |
|
|