Rotor, bucket, tube rack, table for the
5310 model |
|
Rotor Name |
Nominal Capacity |
Tube Size (mm)
diam. x length |
Maximum Speed
(rpm) |
Maximum RCF
(xg) |
Bucket Code No. |
Tube Rack
Adaptor Code No. |
Swinging Bucket Rotor |
ST-720 |
80 x 2mL |
9.5-11 x 36-42 |
3,500 |
2,290 |
053-5820 |
055-4800 |
80
x 10mL |
12-14
x 70-95 |
3,500 |
2,260 |
053-5820 |
055-4740 |
48
x 10mL blood collecting tube/15mL |
15-17.2
x 75-110 |
3,500 |
2,330 |
053-5820 |
055-4750 |
68
x 7mL blood collecting tube |
12.5-13.5
x 75-110 |
3,500 |
2,300 |
053-5840 |
055-0260 |
68
x 10mL blood collecting tube/15mL |
15-17.2
x 75-110 |
3,500 |
2,300 |
053-5840 |
- |
28
x 15mL Conical (TC) |
17
x 121 |
3,500 |
2,300 |
053-5840 |
- |
32 x 15mL Conical (TC) |
17
x 121 |
3,500 |
2,330 |
053-5820 |
055-4780 |
8
x 50mL |
27-36.5
x 75-110 |
3,500 |
2,330 |
053-5820 |
055-4760 |
16
x 50mL Conical (TC) |
30
x 117 |
3,500 |
2,330 |
053-5820 |
055-6070 |
4
x 100mL |
38-46.5
x 75-120 |
3,500 |
2,330 |
053-5820 |
055-4770 |
4
x 175mL Conical (TC)(*1) |
61
x 118 |
3,500 |
2,330 |
053-5820 |
055-4850 |
4
x 200mL Conical (TC)(*1) |
60
x 130 |
3,500 |
2,330 |
053-5820 |
055-4850 |
4
x 225mL Conical (TC)(*1) |
61
x 137 |
3,500 |
2,330 |
053-5820 |
055-4850 |
4
x 250mL |
60-61.8
x 100-130 |
3,500 |
2,380 |
053-5820 |
055-4850 |
ST-410
(*2) |
48
x 2mL |
10.7
x 42 |
4,000 |
2,670 |
Bucket
included |
S17008 |
48
x 5mL |
12.5
x 75 |
4,000 |
2,670 |
Bucket
included |
S17006 |
48
x 7mL |
12-12.5
x 100-112 |
4,000 |
2,670 |
Bucket
included |
S17007 |
28
x 7mL blood collecting tube |
12.5-13
x 75-106 |
4,000 |
2,670 |
Bucket
included |
S17027 |
12 x 15mL Conical (TC)(*3) |
16.5
x 120.6 |
4,000 |
2,670 |
Bucket
included |
S17019 |
12
x 15mL Conical (TC)(*3) |
16.5
x 120.6 |
4,000 |
2,860 |
S13130 |
- |
28
x 10mL blood collecting tube/15mL |
15.5-17
x 90-110 |
4,000 |
2,670 |
Bucket
included |
S17015 |
4
x 50mL |
34
x 110 |
4,000 |
2,670 |
Bucket
included |
S17050 |
4
x 50mL |
27-36.5
x 75-110 |
4,000 |
2,670 |
Bucket
included |
055-5670 |
4 x 50mL Conical (TC) |
29.5
x 117 |
4,000 |
2,670 |
Bucket
included |
S17049 |
4 x 100mL |
44-45
x 110 |
4,000 |
2,670 |
Bucket
included |
S17100 |
4 x 200mL |
56.5
x 104 |
4,000 |
2,670 |
Bucket
included |
- |
ST-480 |
64 x 2mL |
9.5-11 x 36-42 |
3,000 |
1,570 |
053-7030 |
- |
48
x 10mL |
12-14
x 85-110 |
3,000 |
1,600 |
053-7190 |
- |
32
x 10mL blood collecting tube/15mL |
15-17.2
x 85-110 |
3,500 |
2,270 |
053-7130 |
- |
4
x 50mL(*4,*5) |
27-36.5
x 85-110 |
5,000 |
4,640 |
053-7110 |
- |
8
x 50mL(*5) |
27-36.5
x 85-110 |
3,500 |
2,270 |
053-7150 |
- |
16
x 15mL Conical (TC) |
17
x 121 |
3,000 |
1,770 |
053-7300 |
- |
4
x 50mL Conical (TC)(*6) |
30
x 117 |
5,000 |
4,640 |
053-5010 |
- |
8
x 50mL Conical (TC)(*6) |
30
x 117 |
3,500 |
2,270 |
053-5020 |
- |
Microtiter
Plate Rotor |
PT-20 |
6
x Microtiter Plate(*7) |
86(W)
x 128(D) x 52(H) |
4,000 |
1,790 |
Bucket
included |
- |
2
x Deep-well Plate(*7) |
PT-21 |
6
x Microtiter Plate(*7) |
86(W)
x 130(D) x 62(H) |
5,000 |
3,100 |
Bucket
included |
- |
2
x Deep-well Plate(*7) |
PT-89 |
8
x Microtiter Plate(*7) |
86(W)
x 130(D) x 89(H) |
2,400 |
1,080 |
Bucket
included |
055-6370 |
4 x Deep-well Plate(*7) |
Fixed
Angle Rotor |
RA-410M2 |
8
x 15mL Conical (TC) |
17
x 121 |
6,000 |
4,750 |
- |
055-1280 |
8
x 50mL Conical (TC) |
30
x 117 |
6,000 |
4,830 |
- |
- |
RA-500M2 |
10
x 50mL |
26.4-29.2
x 98-121 |
6,000 |
4,830 |
- |
- |
RA-288M2 |
24
x 12mL |
15.9-16.5
x 97-120 |
6,000 |
4,830 |
- |
- |
AT-360 |
24
x 15mL |
15-17.2
x 95-130 |
3,500 |
2,270 |
- |
- |
|
|
(*1) |
175mL Falcon conical
tube (code no. 2076) needs a cushion (code no. 2090).
200mL Nalge Nunc conical tube (code no. 376813) needs
a cushion adapter (code no. 377585).
225mL Falcon conical tube (code no.2075) needs a cushion
(code no. 2090). |
(*2) |
The sealing cap (code
no. S17111) can be used on a rotor of RS-410 series (Separately
sold). |
(*3) |
The sealing cap cannot
be used on this tube rack. The sealing cap (code no. S17111)
can be used on a 12 x 15mL conical (TC) tubes with code
No. S13130. |
(*4) |
The sealing adapter for
50mL conical (TC) (code no. 055-5760) , 15mL conical (TC)
(code no. 055-5750) and 50mL Oak ridge type tubes (code
no. 055-4310) can be used (Option). |
(*5) |
15mL glass tubes can
be used if an adapter (code no. 055-7400) is added and
need a rubber cushion (code no. 024-1400) sold separately. |
(*6) |
15mL conical (TC) tubes
can be used if an adapter (code no. 055-1280) is added. |
(*7) |
If microtiter plates
are used, it is possible to stack them in double or more
on each bucket. However, be sure to test the strength
of those plates prior to centrifugation. The microtiter
plate at the bottom may be crushed. |
|
|