More information on the 3110 model |
RA-528 micro fixed angle rotor
(Option) |
The optional micro fixed
angle rotor, RA-528, accepts various size of micro tubes
from 0.25mL to 2 mL by employing adapters. This rotor
can spin micro tubes over 14,000xg. |
Application Tube
Types |
Nominal
Capacity |
Material
& Shape |
Adapter
Code No. |
0.25mL |
Polyethylene |
055-7580 |
0.4mL |
Polyethylene |
055-7580 |
0.5mL |
Glass tube,
Round bottom |
055-7580 |
0.5mL |
Glass tube,
Tapered |
055-7570 |
0.75mL |
Polyethylene |
055-7590 |
1.0mL |
Glass tube,
Round bottom |
055-7570 |
1.5mL |
Polyethylene |
- |
1.9mL |
Polyethylene |
- |
2.2mL |
Polyethylene |
- |
|
|
|
Automatically changing modes between hematocrit
and 15mL |
In order to improve the
customer's safety, the 3110 model change modes between
hematocrit and 15mL automatically. Hematocrit mode operates
at a very high speed, while 15mL glass mode operates at
low speed. The 3110 allows the customer to change the
mode simply by putting 15mL tubes in the side rack or
putting all 15mL tubes in the rotor. |
|
One-touch removable rotor cover |
The 3110's rotor cover
can save valuable time, enabling installation and removal
by one-touch operation. |
|
Digital display and control knob |
The 3110 model is equipped
with digital display and control knob that provide users
with an easy-to-operate interface. The control knob is
much easier to set each parameter than up-down keys, and
the digital display clearly indicates each parameter. |
|
Interlocking mechanism and twin lock function |
The interlocking mechanism
improves the customer's safety, ensuring completely closing
the chamber lid. Exploiting this system, the 3110 model
will not start to operate even when the [START] key is
pressed. |
|